CT-8000 Battery Testing System
CT-8002S-5V100mA-CV
Portable battery testing system for battery materials research with in-situ testing support. CV Testing, Current response time ≤20μs, Seamless Current Switching, Type-C Power Por, 6-Range Current Measurement
Portable and in-situ testing.
The mini - sized 226×127×48mm device is easy to carry and set up. It supports offline testing and power - failure protection, making it suitable for various complex scenarios:
- Laboratory research:Meets the needs of in - situ battery testing.
- Production line inspection:Facilitates on - site fault diagnosis.
- Outdoor testing:Adapts to mobile testing requirements.
- University teaching:Supports experimental and practical teaching activities.

Depth cyclic voltammetry (CV) testing.
Equipped with CV testing, it can insert CV scans into charge/discharge cycles. Just set the cycle number, and it will automatically perform CV scans to simulate battery aging and verify electrode reaction characteristics after different cycles.

High-precision current control.
With a voltage range of -5V~5V and six current ranges from 0.1μA to 100mA, it achieves ±0.01% of F.S. precision. Using 24bit AD sampling and four - wire connection technology, it ensures accurate capture of tiny signals. The dual - closed - loop control of constant current and voltage provides lab - level precision for battery material research.

Fast response and seamless current switching.
It has a current response time of ≤20μs and can switch currents seamlessly. This allows it to quickly capture dynamic changes during battery charge/discharge, ensuring real - time and accurate data for reproducing true transient conditions.

| Model | CT-8002S-5V100mA-CV |
| Resolution | AD:24bit;DA:16bit |
| Voltage | |
| Output Range | -5V-5V |
| Min discharge voltage | -5V |
| Accuracy | ± 0.05% of F.S. |
| Stability | 0.01% of F.S. |
| Current | |
| Output Range | Range1: 0.1mA |
| Range2: 1mA | |
| Range3: 5mA | |
| Range4: 10mA | |
| Range5: 50mA | |
| Range6: 100mA | |
| Accuracy | ± 0.01% of F.S. |
| Minimum output current | 0.2μA |
| CV cut-off current | Range1: 0.1mA |
| Range2: 1mA | |
| Range3: 5mA | |
| Range4: 10mA | |
| Range5: 50mA | |
| Range6: 0.1mA | |
| Stability | 0.02% of F.S. |
| Power | |
| Single channel output power | 0.5W |
| Accuracy | ±0.02% of F.S. |
| Stability | 0.04% of F.S. |